Online Particle Counter
Model: PT-LPC-O
The LPC-O on-line particle counter is based on the principle of photoresistance (shading) method designated by the International Hydraulic Standards Committee, and is developed according to national and international standards such as GB/T 18854-2002 (ISO11171-1999), which is specially developed for the level of polluted particles in liquids. Detection device. Can provide fast, accurate, reliable and repeatable test results. The operation is simple and convenient, and it is suitable for on-line monitoring of various oil systems. It is also possible to connect two (or more) slaves to the same host or platform for comparison test, with simple operation and flexible combination. It can be widely used in aviation, aerospace, electric power, petroleum, chemical industry, transportation, port, metallurgy, water treatment, machinery, automobile manufacturing and other fields.
Feature
1. Adopt the counting principle of light resistance (shading) method specified by the International Hydraulic Standards Committee
2. High-precision laser sensor, wide test range, stable performance, low noise, high resolution
3. Using high-precision double plunger pump, the sampling speed is stable, and the sampling accuracy is high
4. Built-in pressure reducing device can be used for online high pressure test
5. Built-in common standards such as GJB-420A, GJB-420B, SAE4059E, ISO4406, GB/T14039, NAS1638 and ГOST17216, support custom standard testing, and can set required standards according to customer needs
6. The built-in data analysis system can automatically determine the sample level according to the standard, continuously monitor the hydraulic system and have the alarm function
7. Built-in multiple calibration curves, compatible with all common domestic and foreign standards for calibration
8. Continuous automatic online detection can be realized, and the detection interval can be set arbitrarily
9. With RS232, RS485 and USB interfaces, it can be connected to a computer or laboratory platform, and can also be connected to an industrial field bus
10. With data storage and printing functions, it is convenient for data sorting and filing
11. It can be used as a single machine, and can be used online for upstream and downstream to calculate the filtration rate. It can be monitored by multiple interconnected remote upper computers. It can generate graphs, historical queries, data export, etc., the combination is changeable, and the application scenarios are strong.
12. High-strength metal shell, suitable for use in harsh environments such as on-site
13. The functions and dimensions of the instrument can be customized according to customer requirements
Technical parameters
1. Light source: semiconductor laser
2. Range of particle size: 4μm(c)~500μm(c), 0.8μm~600μm (depending on the selected sensor)
3. Sensitivity: 1μm (ISO4402) or 4μm (c)
4. Sampling volume: 0.2~1000ml (fixed to 10)
5. Sampling volume accuracy: better than ±1%
6. Detection speed: 20mL/min or set according to customer needs
8. Repeatability: RSD<2%
9. Resolution: better than 10%
10. Counting accuracy: better than 10% (less than ±0.1 pollution level)
11. On-line detection pressure: 0.1~0.6Mpa (the maximum pressure of optional pressure reducing device can reach 40MPa)
12. Interval of online detection: 1 second to 23 hours 59 minutes 59 seconds
13. Detection channel: 8 channels arbitrarily set the particle size
14. Protection level: IP56
16. Working temperature: -20℃~60℃
17. Power supply: 100~240VAC or DC24V
Contact: Harry
Phone: 00852-82283663
E-mail: info@lab-kits.com
Add: ROOM 1, 16F, EMPRESS PLAZA, 17-19 CHATHAM ROAD, TSIM SHA TSUI, KOWLOON, HONG KONG