I Series UV-VIS Spectrophotometer

I Series UV-VIS Spectrophotometer
Model: PT-I2/I3/I5/I6/I7/I8/I9

PT-I 2 Visible Spectrophotometer

1.Standard scanning software can directly complete functions of Quantitative; Kinetics; Wavelength Scan; Multi

Wavelength; DNA/Protein and Data processing.

2.Can establish calibration curves and implement associated tests. The instrument internal can be stored with 200 groups of data and 200 standard curves.

3.Suspended posture optical system design, strengthen and thicken the bottom plate to eliminate the vibration or transformations impact on the optical system.

4.Automatic wavelength calibration and automatic deviation repair.

5.Tungsten and Deuterium lamp can be changed easily, without adjustment.

6.Standard with PC software.

 

PT-I 3 UV-VIS Spectrophotometer

1.Standard quantitative software can directly complete photometric analysis, quantitative test and processing of analytical data.

2.Can establish calibration curves and implement associated tests. The instrument internal can be stored with 200 groups of data and 200 standard curves.

3.With calibration curve method, we can establish multiple-point standard curve directly, on basis of which we can measure the concentration of the unknown sample.

4.With coefficient method, we can implement sample measurement directly after inputting coefficient of the curvilinear equation.

5.Automatic wavelength calibration and automatic deviation repair. 6.Deuterium and tungsten lamp can be changed easily, without adjustment. 7.Standard with PC software

 

PT-I 5 UV-VIS Spectrophotometer

1.The main unit and PC software can independently implement functions of Quantitative; Kinetics; Wavelength Scan; Multi Wavelength; DNA/Protein and Data Printing, PC software can complete the function of data processing.

2.Strong function of data processing makes user editing can be easier and more convenient.

3.Suspended posture optical system design, strengthen and thicken the bottom plate to eliminate the vibration or transformations impact on the optical system.

4.Adopt synchronous sine institutions, high accuracy of the wavelength, repeatability.

5.Standard with PC software

 

PT-I 6 UV-VIS Spectrophotometer

1.The Main unit and PC software can independently implement functions of Quantitative; Kinetics; Wavelength Scan; Multi Wavelength; DNA/Protein and Data Printing; PC software can complete the function of data processing.

2.Suspended posture optical system design, strengthen and thicken the bottom plate to eliminate the vibration or transformations impact on the optical system.

3.24-bit high speed and high precision A/D conversion, and improve the sensitivity of the instrument.

4.The core components are imported from Germany and Japan.

 

PT-I 7 UV-VIS Spectrophotometer

1.The Main unit and PC software can independently implement functions of Quantitative; Kinetics; Wavelength Scan; Multi Wavelength; DNA/Protein and Data Printing; PC software can complete the function of data process- ing.

2.Suspended posture optical system design, strengthen and thicken the bottom plate to eliminate the vibration or

transformations impact on the optical system.

3.24-bit high speed and high precision A/D conversion, and improve the sensitivity of the instrument.

4.The core components are imported from Germany and Japan.

5.0.5/1/2/4nm bandwidth can be adjusted automatically.

 

PT-I 8 Double-Beam UV-VIS Spectrophotometer

1.Double beam optical system.

2.The Main unit and PC software can independently implement functions of Quantitative; Kinetics; Wavelength Scan; Multi Wavelength; DNA/Protein and Data Printing, PC software can complete the function of data processing. 3.Suspended posture optical system design, strengthen and thicken the bottom plate to eliminate the vibration or transformations impact on the optical system.

4.24-bit high speed and high precision A/D conversion, and improve the sensitivity of the instrument. 5.The core components are imported from Germany and Japan.

6.The best optical system, based on top structure design, top technological requirements and top raw materials.

7.Standard with PC software

 

PT-I 9 Double-Beam UV-VIS Spectrophotometer

1.Double- beam optical system.

2.The Main unit and PC software can independently implement functions of Quantitative; Kinetics; Wavelength Scan;

Multi Wavelength; DNA/Protein and Data Printing, PC software can complete the function of data processing.

3.Suspended posture optical system design, strengthen and thicken the bottom plate to eliminate the vibration or trans- formations impact on the optical system.

4.24-bit high speed and high precision A/D conversion and improve the sensitivity of the instrument.

5.0.5/1.0/2.0/4.0/5.0 bandwidth can be adjusted automatically

6.The core components are imported with original packaging.

7.The best optical system, based on top structure design, top technological requirements and top raw materials.

8.Standard with PC software

Technical Parameters

Model

PT-I2

PT-I3

PT-I5

PT-I6

PT-I7

PT-I8

PT-I9

Wavelength Range

320-1100nm

190-1100nm

190-1100nm

190-1100nm

190-1100nm

190-1100nm

190-1100nm

Bandwidth

2nm

2nm

1.8nm

1.8nm

0.5/1/2/4nm

1.8nm

0.5/1/2/4/5nm

Wavelength Accuracy

±0.5nm

±1nm

±0.5nm

±0.1nm (D2 656.1nm); ±0.3nm (Full range)

Wavelength Reproducibility

0.2nm

0.3nm

0.2nm

0.1nm

0.1nm

0.1nm

0.1nm

Photometric Accuracy

±0.3%T

±0.5%T

±0.3%T

±0.2%T

±0.2%T

±0.2%T

±0.2%T

Photometric Repeatability

0.15% T

0.2%T

0.15%T

0.15%T

0.15%T

0.15%T

0.15%T

Straylight

0.05%T

0.05%T

0.05%T

0.03%T

0.03%T

0.03%T

0.03%T

Stability

±0.0001A/h(500nm)

±0.001A/h(at 500nm)

±0.0004A/h(at 500nm)

Baseline Flatness

±0.001A/h

±0.0005A

±0.001A

±0.0015A

±0.0015A

±0.001A

±0.001A

Noise

±0.0005A/h

±0.001A

±0.0005A

±0.0005A

±0.0005A

±0.0005A

±0.0005A

Photometric Range

"0-200%T, -0.3--3A, 0-9999C

0-200%T, -4.0--4.0A, 0-9999C

Wavelength setting mode

Automatic

Automatic

Automatic

Automatic

Automatic

Automatic

Automatic

Scanning speed



High, Middle, Low, Optional

Output

USB Port

Printer port

Parallel Port

Display

LCD(320*240)

Light Source

Tungsten Lamp

Deuterium & Tungsten   Halogen Lamp

Detector

Silicon Photodiode

Power

220V AC ±10%/50Hz or 110V AC / 60Hz

Dimension

460x380x180mm

420x300x160mm

460x380x180mm

470x370x180mm

470x370x180mm

625x430x210mm

625x430x210mm

Weight

15kg

13Kg

20Kg

PREVIOUS:Automatic Titrator NEXT:Recirculating Chiller

INQUIRY