Product ID:PT-SX-150
Introduction
The modification of gasoline quality analyzer supports connection with a PC. The octane analyzer can be used in conjunction with a computer.
New oil products grade tester is manufactured on the basis of SX-300 analyzer. The instrument continues the “1XX” series and is designed for testing fuel quality. It has additional modes for
determination of gasoline oxidation breakdown time and kerosene percentage in diesel fuel. The device can be used together with a PC or a laptop through a USB interface. The analyzer supports the automatic computer-based calibration, saving the measurement results in the instrument memory, including the analysis date and time, followed by their transfer to the PC (in Microsoft Excel or txt format). The instrument has an 8-key antiglare keyboard which is easy for operation, wear-proof and resistant to aggressive environments (made in Germany).
A timer microchip is used in the analyzer.
New computer software is included into the set:
• measurement results are reflected on the computer screen in a convenient form;
• the saved data are uploaded from the instrument memory to the PC;
• automatic calibration facilitates the correction.
The instrument is designed for determining the following:
• Determining octane numbers for motor petrols;
• Determining cetane numbers for diesel fuels;
• Determining solidification temperature and diesel fuel type;
• Content of kerosene in diesel fuel;
• Petrol breakdown time (oxidative stability);
The instrument is manufactured as a portable small-sized device and is intended for rapid petroleum products quality monitoring in field and laboratory conditions. Operating conditions: air temperature in the range from - 10 °С up to + 45 °С.
The device is powered from 4 elements of АА type (R6) or from computer USB port.
Instrument Capabilities
Parameter | Description | Compliance |
Measurement of gasoline octane level | Research (RON), motor (MON) method. Anti Knocking Index AKI (Pump octane number PON). | GOST 8226-82 (ASTM D 2699-86), GOST 511-82 (ASTM D 2700-86), GOST Р 51105-97, GOST Р 51866- 2002 (ЕН 228-99) |
Measurement of diesel cetane level | In addition to cetane number, the diesel type and pour point are defined. | GOST 305-82 (ASTM D 4737-03), GOST 3122-67 (ASTM D 613, EN ISO 5165) |
Measurement of diesel fuel pour point | It is determined depending on the diesel fuel cetane number. | Informative parameter |
Determination of diesel fuel type | Diesel fuel type (S-summer, W-winter, and A-arctic) is determined depending on the pour point. | Informative parameter |
Kerosene content in diesel fuel | Kerosene content in diesel fuel is determined as percentage. | Informative parameter |
Gasoline breakdown time (oxidation stability) | Breakdown time is measured in minutes with prior selection of a gasoline grade. | GOST 4039-88 (ASTM D 525), GOST Р 52068-2003 |
Note! The measurement ranges can be customized.
Technical Parameters
# | Parameter description | Measurement units | Value |
1 | Measured petrols octane numbers range | ON | 40-135 |
2 | Acceptable limit of octane number measurement basic error, max | ON | ±0.5 |
3 | Limit of acceptable difference between parallel octane number measurements, max | ON | ±0.2 |
4 | Petrol oxidation breakdown time measurement range | Min | 50-2400 |
5 | Acceptable basic error limit of petrol oxidation breakdown time | min | 10 |
6 | Cetane numbers measurement range | CN | 20-100 |
7 | Acceptable basic error limit of cetane numbers, max | CN | ±1.0 |
8 | Acceptable difference limit between cetane numbers parallel measurements, max | CN | ± 0.5 |
9 | Acceptable error limit when determining diesel-fuel pour point | ºC | 5 |
10 | Kerosene content determination range in diesel fuels | % | 0-95 |
11 | Acceptable basic error limit when determining kerosene content in diesel fuels | % | 3 |
12 | Measurement time | sec | 1–5 |
13 | Insufficient power supply indication operation threshold | V | 5.4 |
14 | Instrument useful life | Minimum years | 6 |
15 | Overall dimensions of: | ||
electronic module | mm | 100х210х40 | |
sensor #1 | mm | 60х100 | |
Instrument weight | g | 680 |
Supplying List
• Electronic computing module 1 piece
• Sensor #1 1 piece
• Sample imitator 1 piece
• Mini-USB cable 1 piece
• (option) Power elements (can be removed) 4 pieces
• User manual 1 piece
• Warranty certificate 1 piece
• Quick start 1 piece
• Tool canvas bag 1 piece