Residual fuel oil total sediment tester

Product ID: PT- SYD-0701

This instrument is designed and manufactured in accordance with the requirements of the People’s Republic of China SH/T 0701 Residual Fuel Oil Total Sediment Determination Method (Hot Filtration Method) and SH/T0702 Residual Fuel Oil Total Sediment Determination Method (Aging Method).

Applies to:

1. In accordance with the requirements of SH/T 0701 standard, determine the total sediment in residual fuel oil with a viscosity of less than 55mm²/s at 100°C and distillate fuel oil blended with residual components.

2. According to the requirements of SH/T 0702 standard, the stability of oil products during storage and transportation can be predicted by measuring the tendency of residual fuel oil to form precipitation.

This instrument is basically equivalent to the technical requirements of ISO 10307-1 and ISO 10307-2 standards, IP375 and IP390 standards.

 

Main technical features

1. This instrument meets the requirements of SH/T 0701 and SH/T 0702 standards. It has two functions in one machine and cost-effective.

2. This instrument is composed of host, steam generator, vacuum pump, constant temperature magnetic stirrer, compact structure, easy to use.

3. The main body of this instrument is built-in aluminum alloy aging bath, copper filter, double heating hole method, uniform heat conduction in the bath.

4. The instrument host has two sets of heating and cooling systems, can be switched according to the test requirements, convenient and fast control.

5. The main body of this instrument has two test holes, which can test two samples at the same time.

6. Standard automatic electric heating steam generator, with mechanical pressure control device, steam output pressure safe and stable. Equipped with built-in and external double safety valve, double safety guarantee.

7. Standard oil-free vacuum pump, without any working medium, easy to use and pollution-free.

8. Standard thermostatic magnetic stirrer, heating temperature and stirring speed can be adjusted.

 

Technical parameters and indicators

(1) Host:

1. Power supply: AC (220±10%) V 50Hz;

2. Aging bath: temperature control range: room temperature120; temperature control accuracy: ±0.5;

3. Heating system: heating power: 600W; heating form: metal bath, two holes;

4. Thermal filter: temperature control range: 10; temperature control accuracy: 1; quantity: two sets;

5. Conditions of use: ambient temperature (1535), relative humidity ≤85%;

6. Dimensions: 570mm×320mm×560mm (length × width × height);

7. Net weight: 15kg;

 

(2) Steam generator:

1. Power supply: AC (220±10%) V 50Hz;

2. Steam pressure: (00.4) MPa;

3. Heating power: 3000W;

 

(3) Vacuum pump:

1. Power supply: AC (220±10%) V 50Hz;

2. Motor power: 160W;

3. Vacuum degree: 200mbar;

 

(4) Constant temperature magnetic stirrer:

1. Power supply: AC (220±10%) V 50Hz;

2. Motor power: 25W; speed: (02400) r/min, (099) electronic gear is adjustable;

3. Heating power: 300W;

4. Digital display temperature control: room temperature99.9.

 