Brief introduction:
full automatic wet laser particle size analyzer adopt MIE scattering principle, measure size is from 0.01μm to 2000 μm, Which offer reliable and repeatable particle size analysis for a diverse range of applications. It use dual-beam& multiple spectral detection systems and side light scatter test technology to significantly improve precision and performance of test, on behalf of the domestic advanced level in the field.
Main Specifications:
Model Name | PSA2008A | PSA2008B | ||
Standard | ISO13320-1:1999, GB/T19077.1-2008, Q/JWN001-2009 | |||
Principle | MIE scattering principle | |||
Measuring Range | 0.01μm-2000μm | 0.01μm-1200μm | ||
Channels Number | 127 | 127 (different photo detectors & optical light path) | ||
Accuracy error | <1% (Deviation of D50 on national standard sample) | |||
Repeatability error | <1% (Deviation of D50 on national standard sample) | |||
Light source | High performance He-Ne Laser (λ= 632.8nm, P>2MW) Lifetime>25000hours | |||
Dispersion Method | Ultrasonic | Frequency:40KHz Power:35W, Time: ≥1S, Anti-dry function | ||
Stir | Revolutions Speed: 0-300RPM (Adjustable) | |||
Circulate | Rated Flow:8L/min Rated Power:10W | |||
Sample Pool | Volume:350mL | |||
Micro- Sample Pool | Volume: 10mL (Available) | |||
Operation Mode | Full automatic/ manual control, freely choose | |||
Resolution | Free distribution truly reflect particle size distribution | |||
Optical bench alignment system | Japan Canon lens, Full automatic, precision is up to 0.1um | |||
Software function | Analysis mode | Free Distribution, R-R Distribution, Logarithm Normal Distribution, Mesh number classification etc. | ||
Statistic Method | Volume Distribution, Quantity Distribution
| |||
Statistic Comparison | Several Testing Results of samples Different batches of samples testing result, Samples before and after processing, Test result of samples in different time.
| |||
User-defined Analysis | Figure out percentage according to the particle size Figure out particle size according to the percentage Figure out percentage according to the particle size range Meet demands of representation of particle test in different industries | |||
Test Report | Word, Excel, Photo( Bmp), Text etc | |||
Multiple-language Support | Multiple language Support | |||
Intelligent operation | Automatically control water inflow, dispersion, test and analysis. Better Repeatability after remove human-factor | |||
Testing speed | <2min/time | |||
Outer dimension | L66cm*W32cm*H40cm | |||
Net weight | 65Kg | |||
Main Features:
1) Patent -Optical path design
Converging light Fourier transform light patented technology, make scattering light not be restricted to lens aperture. Double spectrum design makes its precision achieves from 0.01μm to 2000μm, and Dual-laser orthogonal light make use of the semiconductor auxiliary laser extend the test angle from 45 degree to 135 degree, ensure receive all the angles of signals.
2) Expand Size Range
PSA2008A measurement range significantly expanded to 0.01um, particularly suitable for small samples of the user, the sub-micron particle test results reached the international level
3)Fully automatic built-in wet dispersion system
The ultrasonic dispersion, mechanical stirring and circulation channel reasonable integrated in the interior of the instruments, ensure the uniform dispersion and distribution of particles in the testing process, effectively avoid uneven distribution, large particle deposition phenomenon caused by long pipe line of outside dispersion system, to ensure the accuracy of the test results.
4) Intelligent Operation mode
With intelligent automatic mode of operation, to achieve a key test, as long as according to the prompt addition of sample, click the "test", all processes including water-supply, dispersion, circulation, testing, cleaning, data record, data analysis, save and print are automatically completed, not only reduce the testing workload, but also eliminate the interference of human factors, to further improve the accuracy and authenticity of testing results.
5)Automatic Optical path alignment system
The precision of four hybrid stepping motor in the automatic system of optical components, micro precision of 0.1um, the instrument of light path is always at its best to eliminate manually on the light path and the troubles and difficulties but also enhance the accuracy and stability of test results.
6)Authoritative calibration method
In the entire test range using the national standard materials for calibration, only a year calibration once the instrument can. Calibration method is simple and quick.
7) Unconstrained free fitting patent technique
Original unconstrained free fitting patent technology, particle size analysis is not affected by any functional constraints, truely refelect particle size distribution.
Instrument principle diagram
Adopt Patents Technology:
1.Optical bench design is protected by patent No.- ZL 2014 2 0378380.8,
2.Optical bench alignment system is protected by patent No.- ZL 2013 2 0835882.4.
3.Mie scattering principle application patent No.- ZL 2013 2 0812021.4.
4.Dual laser beam orthogonal application is protected by patent No.-ZL 2007 2 0025702.0
Instrument Test operation interface:
After the background test, click on Energy spectrum test in "test view". And the system will display "test view". Observe the Spectral curve and concentration, After the test result becomes stable, click the “save the results”. The System will automatically save the test result at the present time intervals
Energy Spectrum Test View
Test Report and Its Description:
1) D10: Particle diameter, < 10% particle volume summation percent of total particle’s volume
2) D50: Particle diameter, < 10% particle volume summation percent of total particle’s volume
3) D90: Particle diameter, < 10% particle volume summation percent of total particle’s volume
4) DAV: The average particle size of the particle swarm
5) S/V: Volume specific surface area; surface area per unit volume of particles
Particle size analysis diagram:
l Horizontal axis is the size of the particle, which is the logarithm distribution.
l Left column is the volume of the cumulative percentage, corresponding to the upward trend curve
l Right column is the volume percentage of a certain interval, corresponding to the histogram or the ups and downs of the curve.
l The test result of data list is corresponding to the analysis chart.
Application:
PSA2008 is widely used in mineral materials, cement, ceramics, chemicals, paint, emulsion, dyes, pigments, fillers, chemical products, catalysts, drilling mud, abrasive, lubricant, coal, sand, dust, bacteria, cells, food, additives, pesticides, explosives, graphite, photosensitive materials, fuel, and ink metal and non-metal powder, calcium carbonate, kaolin, coal water slurry and other powder material.
Contact: Harry
Phone: 00852-82283663
E-mail: info@lab-kits.com
Add: ROOM 1, 16F, EMPRESS PLAZA, 17-19 CHATHAM ROAD, TSIM SHA TSUI, KOWLOON, HONG KONG