Laboratory Electrochemical Electrodes
Glass Membrane
Glass membrane is the most important part of pH electrode.
● LabSen pH electrodes are equipped with 4 types of glass membrane to meet various
applications: S membrane, H membrane, HF membrane and PHY membrane.
For example HF membrane is used for HF resistance electrode.
● LabSen glass membrane has good impact resistance. It will not be damaged by general
intensity impact, completely differentiating from conventional glass membranes.
● LabSen glass membrane with different shapes are shown as below:
Junction
Junction is the electrolyte interface between reference system and the solution to be
measured, LabSen electrode adopts the following types of junctions:
● Ceramic -- the most frequently used junction, easy to be blocked by protein-containing
or suspension solution.
● Pore without diaphragm -- it is used with solid electrolyte, no clogging, maintenance-free.
● Movable sleeve -- easy to clean, suitable for suspension, emulsion, low ion concentration
solution and nonaqueous solution. The infiltration rate of electrolyte is determined by the
tightness of the sleeve during installation.
● PTFE -- a kind of Teflon material with multi pores, hard to be contaminated.
Inner Solution
The inner solution of LabSen electrode is in a unique dark blue. With a special gel
treatment, the inner solution does not flow and will not cause air bubbles. The electrode can
work well even when being upside down.
Reference System
Besides routine Ag/AgCl reference electrode, LabSen pH electrodes are more likely to
adopt Long-life reference electrode and Silver-ion-trap reference electrode.
Long-life reference system is composed of a glass tube, AgCl and reference silver wire.
The top end of the slim glass tube is stuffed with cotton, which will prevent reaction between
AgCl and electrolyte when temperature changes, improving the stability of reference
electrode and service life.
Based on the Long-life reference system, a Silver-ion-trap is added to prevent the contamination of the junction when testing
samples contains sulfides and proteins, significantly increasing the stability of the reference electrode.
Electrolyte
In addition to the conventional 3M KCl solution and gel type, LabSen pH electrode adopts the following special electrolyte:
● Polymer: a new type of polymer reference electrolyte. Main technical features: The reference electrode is with single pore or
multi-pore, without diaphragm. The polymer electrolyte directly contacts with test samples. The anti-pressure capacity
can reach 6 bars. It is suitable for almost all applications, including samples containing oil, low ion concentration, protein,
and suspension.
● Protelyte: suitable for testing samples containing protein, low temperature samples and viscous samples
Swiss patented Technology- pH/Temp. Combination Structure
The LabSen pH electrode adopts Swiss patented pH/Temp. combination structure. The conventional way to make the
glass pH/Temperature combination electrode is to place the temperature sensor inside the bulb or electrode. As the bulb and the electrodes are filled with solution, the temperature sensing is very slow. Adopting the patented structure, the temperature sensor is placed in an independent cavity under the pH glass membrane. With built-in heat conducting medium, the temperature sensing speed can be increased by 40%. The response rate and stability of the pH electrode are greatly improved.
Double Junction pH Electrode for Ultrapure Water
The following picture is the sketch of LabSen pH electrode for ultra-pure water. There are two junctions and two reference cavities.
It is Junction I in movable sleeve form that will be in contact with ultrapure water. The reference solution is 1M KCl. The density
gradient and polarization effect on the glass membrane will be decreased and the measuring stability will be greatly improved.
pH Electrode with Pre-pressure
The Pre-pressurized electrode is an electrode prefilled with certain pressure. Even in highly viscous samples, the electrolyte can still infiltrate well, preventing from clogging, ensuring the stability and repeatability of measurement. This type of electrode is especially suitable for testing highly viscous samples, like cosmetics, paint, resin, etc.
Technical parameters
Model | LabSen211 | LabSen213 |
Name | Routine pH Electrode | Routine 3-in-1 pH Electrode |
Picture | ||
Application | routine laboratory use | routine laboratory use |
Features | impact-resistant glass membrane | impact-resistant glass membrane, patented ATC structure. |
Body | Lead-free Glass | Lead-free Glass |
Temp. Probe | No | NTC 30kΩ |
Junction | Ceramic | Ceramic |
Reference | Long-life | Long-life |
Electrolyte | 3M KCl | 3M KCl |
Length | 120mm | 120mm |
Diameter | 12mm | 12mm |
Membrane Shape | Hemispherical | Cylindrical |
Range | 0~14 pH | 0~14 pH |
Temperature | -5~100℃ | -5~100℃ |
Connector | BNC/1m cable | BNC/RCA 1m cable |
Model | LabSen231 | LabSen221 |
Name | Premium pH Electrode | Premium pH Electrode |
Picture | ||
Application | suitable for wastewater, emulsion, suspension, effluent. | suitable for viscous and low ion concentration samples. |
Features | with 2-pore and solid electrolyte, hard to be contaminated, high compression-resistance. | with movable sleeve and BNC connector. |
Body | Lead-free Glass | Lead-free Glass |
Temp. Probe | No | No |
Junction | 2-pore | Movable Sleeve |
Reference | Long-life | Long-life |
Electrolyte | Polymer | 3M KCl |
Length | 120mm | 130mm |
Diameter | 12mm | 12mm |
Membrane Shape | Hemispherical | Cylindrical |
Range | 0~14 pH | 0~14 pH |
Temperature | -5~80℃ | -5~80℃ |
Connector | BNC/1m cable | BNC/1m cable |
Model | LabSen223 | LabSen331 |
Name | Precise pH Electrode | Plastic Premium pH Electrode |
Picture | ||
Application | suitable for viscous and low ion concentration samples, integrated ATC structure. | suitable for wastewater, emulsion,suspension, effluent |
Features | with movable sleeve and BNC/RCA | With anti-corrosion body, 1-pore, solid |
Body | Lead-free Glass | POM |
Temp. Probe | NTC 30kΩ | No |
Junction | Movable Sleeve | 1-pore |
Reference | Long-life | Long-life |
Electrolyte | 3M KCl | Polymer |
Length | 130mm | 120mm |
Diameter | 12mm | 12mm |
Membrane Shape | Cylindrical | Spherical |
Range | 0~14 pH | 0~14 pH |
Temperature | -5~80 |
Contact: Harry