UV Visible Spectrophotometer
Model ID:ST-UV-1300PC, ST-UV-1500PC
Character
●Single-chip microcomputer control, LCD Screen (128*64 Dots)
●Large LCD display can display multiple sets of data
●Huge memory space to store multiple sets of data and curves
●Auto zero,100% adjustment function
●Wavelength automatic adjustment
●Wrap lights, automatic filter change, large sample pool (0.5mm ~ 100mm)
●Standard samples having a maximum ten-point standard curve measurement functions built
●Direct input by K, B factor for quantitative measurement standard curve
●Can be directly input and the corresponding standard concentration values establish a standard curve for quantitative measurement
●Adjustable electric saving standard curve parameter measuring set, can save a plurality of standard curve
●Can be edited at any time of the standard curve, user-friendly
●General Parallel printer interface with or slightly beat interfaces can print the title bar, the measurement data, curve parameters, the standard curve and the curve sample points
●With USB interface
●PC control can be achieved through more accurate and flexible measurement requirements (standard)
Software features
●Wavelength Scan: the measured object can be 190nm-1100nm full band spectrum scan, can also be segmented scan
●Kinetic tests: the time of the measured object scanning, timely detection of changes in concentration caused by changes in absorbance
●Quantitative measurement: by establishing a standard curve, the concentration can be measured in the analysis was
●Multi- wavelength testing: set a plurality of wavelengths simultaneously, the measured absorbance of a plurality of wavelengths in the test
●DNA / protein test: the test formula purity of nucleic acid and protein concentration analysis of DNA / protein
Technical Specifications
Model | ST-UV-1300PC | ST-UV-1500PC |
Wavelength Range | 190~1100nm | |
Band Width | 4nm | 2nm |
Wavelength Accuracy | ±0.5nm | |
Wavelength Repeatability | ≤0.2nm | |
Photometric Accuracy | ±0.5% τ | |
Photometric Repeatability | 0.2% τ | |
Stray Light | ≤0.05% τ (220nm NaI, 340nm NaNO2) | |
Stability | 0.001A/h@500nm | |
Photometric Mode | T, A, C, E | |
Wavelength Setting | Automatic | |
Photometric Display Range | -0.3~3A | |
Display Mode | LCD Screen (128*64 Dots) | |
Detector | Import Silicon Photodiode | |
Light Source | Import Deuterium Lamp & Tungsten Halogen Lamp | |
Power Requirement | AC 220V/50Hz 110V/60Hz | |
Power | 120W | |
Dimensions(W*D*H) | 480×350×220mm | |
Weight | 15Kg |
Standard configuration
Spectrophotometer main unit 1set
1cm quartz cuvette 2pcs
1cm glass cuvette 4pcs
Dust cover 1pc
Power line 1pc
Instruction manual 1pc
Software suite 1set (Standard)